Courtney et al. (2008) carried out a multi-centre, single-arm, prospective case series evaluating treatment outcomes in 34 patients with malignant ascites treated with the PleurX peritoneal catheter drainage system over a 12‑week follow-up period (or until death in some patients). It reported 100% technical success during the placement procedure (defined as intraperitoneal positioning of the device and the ability to withdraw ascitic fluid from the device at the completion of the procedure), with one minor procedural complication. Twenty patients experienced complications during the follow-up period including minor complications that resolved spontaneously. Two catheters needed to be removed, and other complications were infection (n=2), occlusion/loculations (n=4), leakage of ascitic fluid (n=7), dizziness (n=5), shortness of breath (n=1) and severe anaemia (n=1). Available records from 19 patients showed that the mean number of drainage sessions after placement of the PleurX peritoneal catheter was 23.3 per patient (range 5 to 56), and that of the total 433 sessions, 13% were performed by a nurse, and the remainder were carried out by the patient alone (28%) or a carer (58%). The catheter patency rate was 85% (n=29); the remaining 5 patients were lost to follow-up. Changes in symptom severity between baseline and at 2, 8 and 12 weeks were assessed using a validated tool. There was a reduction in the severity of abdominal discomfort, bloating, diarrhoea and nausea at 2 weeks and 8 weeks. An overall improvement in quality of life at 12 weeks was reported in 28% of respondents.